Buying an Electrosurgical Generator: A Guide for Surgery Centers and Clinics

An electrosurgical unit (ESU) is one of the most-used pieces of equipment in an operating room, and one of the easiest to buy badly. Generators look similar from the front panel, but they differ in output modes, in how they protect the patient at the return electrode, and in which handpieces, pads and smoke evacuators work with them. This guide covers what to settle before you ask for quotes, if you are equipping a new OR, replacing an older unit or adding a hyfrecator to a dermatology room.

Match the unit to the procedures

Start with the cases the generator will support.

Output modes

Generators are described by the modes they offer, and the names vary by manufacturer. Medtronic's operator's guide for the Force FX-C lists three bipolar modes (precise, standard and macrobipolar), three monopolar cut modes (low, pure and blend) and three monopolar coagulation modes (desiccate, fulgurate and spray). It also supports simultaneous coagulation from two handpieces, activation by hand or footswitch, and use with the Force Argon system for argon-enhanced coagulation.

When comparing units, ask your surgeons which modes they actually use and make sure the generator offers them. A unit with fewer modes is not a problem if it covers your cases.

Return electrode monitoring

In monopolar surgery, current returns through a pad on the patient. If the pad lifts or loses contact, the current density under it can rise and burn the patient. Most modern OR generators include a contact-quality monitoring system for this. Valleylab calls it REM. According to Valleylab's service manual for the Force FX-C, when a split REM pad is used, the generator measures the impedance between the two pad halves. It alarms and stops output if that impedance goes outside 5 to 135 ohms or rises more than 40 percent above the baseline taken at application. With a non-REM pad, it only checks that the cord's resistance is low. That is why the pad type and the REM return-electrode cord matter as much as the generator itself.

Return electrodes are also worth watching for recalls and safety notices. Before you standardize on a pad or reusable return-electrode system, check the manufacturer's current notices.

Accessories and compatibility

The generator is only part of the system. Confirm:

  • Pencils and handpieces. Connector types differ between brands and generations. A reusable pencil such as the Valleylab E2100 needs a compatible socket.
  • Footswitches. Monopolar and bipolar footswitches are separate items, such as the Valleylab bipolar footswitch.
  • Return electrodes and cords, as covered above.
  • A cart, such as the Valleylab generator cart, if the unit will move between rooms.

Plan for surgical smoke

Electrosurgery produces smoke. NIOSH's hazard control guidance on surgical smoke notes that it can contain toxic gases and vapors such as benzene, hydrogen cyanide and formaldehyde, as well as particles. NIOSH recommends local exhaust ventilation with a smoke evacuator. It suggests keeping the capture nozzle within 2 inches of the surgical site, running the evacuator whenever smoke is produced, and using new filters and tubing for each procedure. It also notes that room suction systems are much less effective at capturing smoke.

If you are buying a generator, it is a good time to price a smoke evacuator alongside it, such as the Bovie Smoke Shark or Buffalo Filter Porta PlumeSafe 604. Some state rules and facility policies now require smoke evacuation, so check what applies to you.

Buying a refurbished ESU

Most of the generators in our ESU range are refurbished, which can make good sense for a surgery center or a clinic adding rooms. Ask the supplier for:

  • The serial number, software version and date of manufacture.
  • A record of the output power checks and electrical safety checks done.
  • Confirmation that return-electrode monitoring and alarms were tested.
  • Which accessories are included, and what you will need to buy separately.
  • Warranty terms, and whether the manufacturer still supports the model with parts and service.

Our refurbished equipment checklist covers the general questions. Have your biomedical engineering team or service provider do an incoming inspection before first use.

Frequently asked questions

What is the difference between an electrosurgical generator and a hyfrecator?

A hyfrecator is a low-power unit for desiccating and fulgurating superficial tissue, mainly in dermatology. A full electrosurgical generator offers higher power, cutting modes and bipolar output for surgery in the OR.

Do I need a generator with return electrode monitoring?

For monopolar surgery in an OR, a contact-quality monitoring system such as Valleylab's REM is a key safety feature. It is standard on most modern OR generators. Use pads that work with it.

Is a smoke evacuator required?

NIOSH recommends local exhaust ventilation with a smoke evacuator. Requirements depend on your state and facility policy, so check what applies to you.

Next steps

Browse our electrosurgical generators and ESU accessories, or request a quote with the models you are considering, the procedures, the accessories you need and the delivery location.

Need a quotation for equipment?

Send us the equipment you need, the quantity and where it should be delivered. We will review your request and respond with available options, pricing and shipping information.